Sfoglia online o scarica il pdf Manuale di istruzioni per Generatore portatile Com-Power CG-515. Com-Power CG-515 11. Comb generators

Comb Generators models CG‐515 
 

Product Description

Overview 
 
The CG‐515 Comb Generators are reference signal sources.  They were designed for EMC laboratories use. It is 
ideal for quickly detecting malfunctioning equipment in the radiated emissions test setup in between full 
calibration intervals.  These malfunctions may go undetected, unless it they are checked often.  Full site calibration 
although more accurate is time consuming is not practical for a quick test.  By comparing Comb Generator data 
taken at regularly, the EMI lab can ensure that all the test equipment and accessories are operating within the 
expected range. If any abnormality is detected, it can be quickly investigated and rectified before it causes 
measurement errors.  
 
The Comb Generators can also used for comparing two or more sites with similar setup. They can also be utilized 
as conducted signal source for testing Line Impedance Stabilization Network (LISN), filters and cables.  
 
Items included with Comb Generators 
 
Equipment, accessories, and documents supplied with the model CG‐ 515 Comb Generators are as follows: 