According to the prevailing operating conditions and requirements, the sensor may require regu-
lar recalibration to ensure proper functionality. The relevant intervals for recalibration, if any,
must be defined by the user according to the very individual requirements.
There are different options for sensor calibration:
Sensor calibration at Baumer (return sensor for maintenance)
n
Sensor calibration with the aid of a reference sensor
n
Sensor calibration with the aid of a reference medium
n
Sensor calibration using the Baumer calibration box (dry calibration)
n
11.3.1
Sensor calibration with the aid of a reference sensor
In this method, the medium values for conductivity and temperature are measured both by an
appropriate reference sensor and the sensor which is to be calibrated. Deviations in the two
measurement results are saved as offset in the sensor to be calibrated.
INFO
This task can be done either using the FlexProgram or via IO-Link (e.g. BSS). The following ex-
ample shows calibration with FlexProgram.
1
Reference sensor
3
FlexProgrammer
Condition:
ð The sensor is connected to a PC with software FlexProgram installed.
ð The sensor and the reference sensor are mounted in one medium.
Instruction:
a) Open software FlexProgram on the PC.
b) Open setting PAC50 in the left column.
c) Click on icon
d) Click on button
ü
Dialog window opens.
e) Navigate to
V1 | PAC50H/S
1
Measuring
(Online).
Calibrate Sensor
below the visualization.
Calibrate
offset.