Model 100A250A
2.2.2
2.2.2.4.2 IEEE-488 device address selection (continued)
Device Address
2.2.2.5
Command Set Format
Each command is composed of one alpha character, up to four numeric parameters, and a
command termination character. The command termination character is the "line feed" command,
which is denoted and entered as <LF>. Commands are case-sensitive and must be entered in upper
case only in order to be recognized.
2.2.2.6
IEEE-488 Communications
For IEEE-488 communications, the "End or Identify" (EOI) control line may also be used for
command termination. When sending commands to the Model 100A250A via the IEEE-488 bus,
terminate each command with a <LF>, an EOI, or both. No characters are permitted after the
<LF> or EOI; the 100A250Ainterprets characters following the <LF> or EOI as the start of the
next command. When an error condition is present at the Model 100A250A, the "Service Request"
(SRQ) line is asserted; the operator can then perform a serial poll operation. The Model 100A250A
error code (in binary) is contained in the returned serial poll status byte (STB). These error codes
are defined in Table 2-2.
Remote Operation (continued)
IEEE-488 Device Address Selection
Switch 5
1
off (0)
2
off (0)
3
off (0)
4
off (0)
5
off (0)
:
:
30
on (1)
Table 2-1
Switch 4
Switch 3
off (0)
off (0)
off (0)
off (0)
off (0)
off (0)
off (0)
on (1)
off (0)
on (1)
on (1)
on (1)
Manual Text
Switch 2
Switch 1
off (0)
on (1)
on (1)
off (0)
on (1)
on (1)
off (0)
off (0)
off (0)
on (1)
on (1)
off (0)
2-4