Follett Edge Series Manuale di installazione - Pagina 2

Sfoglia online o scarica il pdf Manuale di installazione per Frigorifero Follett Edge Series. Follett Edge Series 17. Laboratory freezer
Anche per Follett Edge Series: Manuale di installazione (17 pagine)

Follett Edge Series Manuale di installazione

Preliminary Information

Caution Flammable Refrigerant
Caution: Risk of fire.
§ Risk of fire or explosion. Flammable refrigerant used.
Before attempting to install or service this product, all safety precautions must be followed.
CAUTION! – Risk of fire or explosion. Dispose of properly in accordance with federal or local regulations.
Flammable refrigerant used.
CAUTION! – Risk of fire or explosion due to puncture of refrigerant tubing; follow handling instructions carefully.
Flammable refrigerant used.
CAUTION! – Handling, moving, and operating the freezer to avoid either damaging the refrigerant tubing or
increasing the risk of a leak.
§ If the equipment is not transported vertically, a period of at least 24 hours must be allowed before starting.
The equipment must be transported and handled exclusively in upright position. This is required to avoid
contamination of the refrigerant with compressor oil. Contamination could result in valve and heat exchanger coil
failure, problems starting the electric motor, or the risk of gas leak. The manufacturer is not responsible for any
problems due to transport executed in conditions other than those specified.
The equipment must be handled using a fork lift truck or a pallet truck with suitable forks (fork length at least equal
to 2/3 length of unit).
§ Read this manual thoroughly before operating, installing or performing maintenance on the equipment. Failure to follow
instructions in this manual can cause property damage, injury or death.
§ The machine contains R600a, refrigerant. R600a (isobutane) is flammable.
§ When servicing this equipment, be sure to lock the circuit breaker, and display an in-service notice.
§ This equipment contains high-voltage electricity and refrigerant charge. Installation and repairs are to be performed by
properly trained technicians aware of the dangers of dealing with high voltage electricity and refrigerant under pressure. The
technician must also be certified in proper refrigerant handling and servicing procedures. All lockout and tag out procedures
must be followed when working on this equipment.
§ Repair on R600a systems must always be done in a well-ventilated area.
§ Because R600a is highly flammable, a combustible gas leak detector is required when servicing R600a systems.
§ Only use parts recommended or provided by the manufacturer.
§ Do not tilt unit further than 30° off vertical during uncrating or installation.
§ If the supply cord is damaged, it must be replaced by the manufacturer, its service agent or similarly qualified persons in
order to avoid a hazard.
§ This appliance is designed for commercial use.
§ To reduce risk of shock, disconnect power before servicing.

General Information

§ To avoid temperature rise in the cabinet and/or contents, care should be taken to ensure that the door is
opened only for brief periods.
§ Ensure that the intake and exhaust vents are not blocked. Adequate space should be provided between the
items inside the unit to allow air circulation.
§ In the freezer compartment, put stored items on the shelves and do not let them contact the wall.
§ Fix the shelves securely. Place items on the shelves and leave a space between the wall of the cabinet and
the contents to allow air circulation. Do not place items on the floor of the chamber.
§ The unit shall not be exposed to direct sunlight and shall not be near a heat source such as a heater, stove,
or other heat dissipating equipment (sterilizer, autoclave, etc.).
2
FZRNAT5-SD
DANGER
CAUTION!
WARNING!