Operating Instructions
The table below shows the basic operation method. Perform key operation in the sequence indicated in the table.
The example in the table assumes that the refrigerator temperature is 6.0 C.
Note: The unit is set at the factory at 5.0 C.
Basic Operation Procedure (Example of setting the refrigerator to 4.0 C)
Operation
1
Connect to the power source.
2
Press SET key.
3
Set to 4 by using the down
arrow key.
4
Press and release SET key.
Note: If no key has been pressed for 90 seconds in the temperature set mode, the display mode returns
automatically to the temperature display mode. In this case, the chamber temperature setting is not
changed. The refrigerator temperature can be set in the range between 2.0 C and 8.0 C. The guaranteed
temperature with no load at an ambient temperature of 22 C is 2 C to 8 C.
Key operated
-----
The current refrigerator temperature is
displayed after 5 seconds of connecting the
unit to the power source. When refrigerator
temperature is higher than 25 C, HI is
displayed and LO is displayed when lower
than –10 C.
SET
The digits of the temperature display flash.
Pressing and releasing the down arrow key
shifts down the set point by 0.1 C increments.
SET
The value (4.0) is stored in memory and the
current refrigerator temperature is displayed.