Amprobe 30XR-A Manuale d'uso - Pagina 8

Sfoglia online o scarica il pdf Manuale d'uso per Multimetro Amprobe 30XR-A. Amprobe 30XR-A 14. Professional digital multimeter
Anche per Amprobe 30XR-A: Manuale d'uso (14 pagine)

Amprobe 30XR-A Manuale d'uso
Specifications

General Specifications

Display: 3½ digit liquid crystal display
(LCD) with a maximum reading of
1999.
Polarity: Automatic, positive implied,
negative polarity indication.
Overrange: (OL) or (-OL) is displayed.
Zero: Automatic.
Low battery indication: The B is
displayed when the battery voltage
drops below the operating level.
Measurement rate: 2.5 per second,
nominal.
Operating environment: 0 °C to 50 °C
at < 70 % R.H.
Storage temperature: -20 °C to 60 °C,
0 to 80 % R.H. with battery removed
from meter.
Temperature Coefficient: 0.1 ×
(specified accuracy) per °C. (0 °C to
18 °C, 28 °C to 50 °C).
Environment: Indoor use, Altitude up to
2000 m
Power: Single standard 9-volt battery,
NEDA 1604, JIS 006P, IEC 6F22.
Battery life: 200 hours typical with
carbon-zinc.
Dimensions: 196 x 92 x 60 mm ( 7.7" x
3.5" x 2.4").
Weight: Approx. 426 .( 0.94 lb.) without
holster, including battery.
Accessories: One pair test leads
(TL36), 9 V battery (installed), Magna
Grip™ Holster, and Operating
Instructions.
Warranty: One (1) Year
Approvals:
>
LISTED
950Z
Safety: Conforms to UL1244;
EN61010-1: Cat II - 600V / Cat III -
300V; Class 2, Pollution degree II.
The 30XR-A is recommended for use
with local level power distribution,
appliances, portable equipment, etc,
where only smaller transient
overvoltages may occur, and not for
primary supply lines, overhead lines
and cable systems.
EMC: Conforms to EN61326-1.
This product complies with
requirements of the following European
Community Directives: 89/ 336/ EEC
(Electromagnetic Compatibility) and 73/
23/ EEC (Low Voltage) as amended by
93/ 68/ EEC (CE Marking). However,
electrical noise or intense
electromagnetic fields in the vicinity of
the equipment may disturb the
measurement circuit. Measuring
instruments will also respond to
unwanted signals that may be present
within the measurement circuit. Users
should exercise care and take
appropriate precautions to avoid
misleading results when making
measurements in the presence of
electronic interference.
Electrical Specifications
(at 23 °C ± 5 °C, <75 % R.H. non-
condensing)
DC VOLTS
Ranges: 200 mV, 2 V, 20 , 200 V, 600 V
Accuracy: All ranges, ± (1.0 % rdg +
1 dgt)
Resolution: 100 μV in 200 mV range
Input impedance: 10 MΩ
Overload protection: 200 mV range:
600 V dc or 600 V ac rms 15 seconds.
Other ranges: 600 V dc or 600 V ac rms
7
P