Amprobe ac68C Manuale d'uso - Pagina 9
Sfoglia online o scarica il pdf Manuale d'uso per Multimetro Amprobe ac68C. Amprobe ac68C 14. True rms ac/dc clamp multimeter
Resistance Measurement
Before taking any in-circuit resistance measurement, remove
power from the circuit being tested and discharge all
capacitors.
1.
Set the function/range selector to the 4 kΩ position.
2.
Connect the black lead to the COM terminal and red lead to
the "VΩ" terminal.
3.
Connect the test leads to the resistance or circuit to be
measured and read the value on the display.
Continuity Measurement
Before taking any in-circuit continuity measurement, remove
power from the circuit being tested and discharge all
capacitors.
Set the function/range selector to the R position.
1.
2.
Connect the black lead to the COM terminal and red lead to
the "VΩ" terminal.
3.
Connect the test leads to the circuit to be tested.
Continuity (resistance ≤40 Ω) is indicated by a continuous
beep tone.
Diode Test
Set the function/range selector to the G position.
1.
2.
Connect the black lead to the COM terminal and red lead to
the "VΩ" terminal.
3.
Connect the red test lead to the anode, and the black test
lead to the cathode of the diode to be tested.
4.
Read the forward voltage value on the display (approx.
0.6 V for a silicon diode or 0.4 V for a germanium diode).
An open diode is indicated by "OL".
5.
Reverse test lead connections to the diode to perform a
reverse bias test. "OL" indicates a good diode.
Notes: "OL" for both reverse and forward bias tests indicates an
open diode. A low voltage reading for both bias tests indicates a
shorted diode. If the diode is shunted by a resistor of 1000 ohms
or less, it must be removed from the circuit before taking the
measurement. Bipolar transistor junctions may be tested in the
same manner described above as emitter-base and base-
collector junctions are diode junctions.
Peak Hold Measurement
Proceed as for voltage or current measurement (AC or DC).
Before actually taking the measurement, press the PEAK button.
"PEAK" is shown on the display. After taking a measurement, the
displayed reading is the maximum peak value of a surge in the
voltage or current.
W
Warning
W
Warning
5