Folyon H300P Manuale d'uso - Pagina 11

Sfoglia online o scarica il pdf Manuale d'uso per Camera climatica Folyon H300P. Folyon H300P 17. Temperature humidity chamber with watlow pm plus controllers

Press the Home key when done.
To turn the control back On follow these steps:
Press the Forward Arrow key once. Control Mode 1 should be highlighted.
Press the Forward Arrow key once.
Press the + or – key until Auto is highlighted.
Press the Home key when done.
Setting Up a Profile (PM6R Models Only)
For instructions on how to set a profile, see the "PM Plus User's Manual", Chapter 6: Profiles. For easy
profile setup download the Watlow EZ-LINK™ App to your phone or tablet. The App communicates with the
controllers using Bluetooth®.

Setup Page Parameters

The setup page parameters below have been programmed into the temperature and humidity controllers at
the factory. These parameters are for reference only and should not be modified.
Improper modifications to these setup values can result in erratic performance and unreliable operation. Do
not attempt to modify the setup values, unless you thoroughly understand what you are doing.
These parameters were saved at the factory under User Settings 1. If these parameters are modified, they
can be restored to factory defaults. Please see the Restoring Factory Default Parameters section in this
manual on how to do this.

Analog Input Menu

Parameter
Sensor Type
RTD Leads
Units
Scale Low
Scale High
Range Low
Range High
Process Error Enable
Filter
Input Error Latching
Display Precision
Calibration Offset
H300P Temperature Humidity Chamber
Temperature
Humidity
RTD 100 Ohm
Milliamps
3
Relative Humidity
4.00
20.00
0.0
100.0
Off
0.5
0.5
Off
Off
Tenths
Tenths
0.0
0.0
Page 11