2.1 High resistance test

This test is only carried out with pH instruments. The high electric resisitance of the pH
glass electrode requires an equally high resistance wihtin the whole measuring
system. If the reistance drops, for e. g. due to humidity, the measurement becomes
unstable and tends to drift or wave.
To test the high resistance of the instrument and the cable, the PH-T is equipped with
an even higher internal resistance that can be activated with the key labelled "MegaΩ".
If the resistance is high enough throughout your measuring system, the activation has
no effect - the values displayed by the instrument and the PH-T remain more or less
unchanged. If not, then the value will decrease drastically.
1 - Connect
Connect the PH-T via the adapter to the pH cable of your instrument.
2 - Switch on
Switch on the PH-T by pressing key "ON" for 5 seconds.
NOTE
Switch-on is delayed to prevent accidental activation during transport.
3 - Adjust value
Stay in the mV mode. Adjust the value to approx. 300mV. Your pH instrument should
show now approx. pH 2.
4 - High resistance test
Press the key labelled "MegaΩ". "MΩ" in the display indicates that the high resistance
has been activated.
Focus on the value displayed by the PH-T: if it drops significantly, e. g. to approx. half its
original value, then the measuring system does not provide the required high
resistance.
NOTE
Often the cable rather than the instrument is the source of loss in
resistance - to make sure, repeat the test with a different cable!
NOTE
The high resistance test consumes a lot of battery power - don´t
forget to switch it off by pushing again the key "MegaΩ Ω Ω Ω Ω "!