M620 Series
Common Issues (Continued)
Symptom
Cause
The color of the image is
The brightness adjustment knob is not in the right position
not accurate
No filter is used or filter is in use
The objective touches the
The cover slip is not standard sized
cover slip
The limit-stop is set too high or not engaged
Unable to move the slide
The slide is not secured correctly
smoothly
The mechanical stage is not properly secured
Focus knob does not turn
The tension knob is too tight
Stage declines by itself
The tension knob is too lose
The focusing knob won't
Limit-stop is engaged
raise the stage
The bulb/light source
The bulb is close to burning out
flickers
The microscope does not
The microscope is unplugged
light up
Remedy
OPTICAL ISSUES
Adjust the brightness knob to a higher or lower setting for color
clarity
Remove color filter if natural light is desired, or insert desired
filter
MECHANICAL ISSUES
Replace the cover slip with the appropriate 0.17mm thickness slip
Be careful to avoid contact between objective and the slide when
the limit stop is not engaged. To reengage, focus the sample, then
lock the limit stop into place to set max height at a safe but us-
able distance.
Adjust the slide to use the stage clips and secure the sample
Tighten the mechanical stage screws to better secure the stage
Loosen it by adjusting the tension ring inside the coarse focus
knob counterclockwise (close to the arm of the microscope on
the left of the microscope)
Tighten it by adjusting the tension ring inside the coarse focus
knob clockwise (close to the arm of the microscope on the left
of the microscope)
Disengage the limit stop on the rear of the stage of the micro-
scope
ELECTRICAL ISSUES
Please contact us in regards to this issue. LED lights do not burn
out, so there may be another electrical issue
Insert the plug into the wall socket to achieve electrical illumina-
tion
17
Troubleshooting