Dr. Honle Dermalight 80 UV-B-311 nm Manuale di istruzioni per l'uso - Pagina 12

Sfoglia online o scarica il pdf Manuale di istruzioni per l'uso per Apparecchiature mediche Dr. Honle Dermalight 80 UV-B-311 nm. Dr. Honle Dermalight 80 UV-B-311 nm 17.

Dr. Honle Dermalight 80 UV-B-311 nm Manuale di istruzioni per l'uso
M
8.
AINTENANCE
Daily:
Check the filter glass for damage and dirt; check the housing for
damage.
Every 300 operation hours :
Replace the lamps.
Every 2 years: Maintenance by our Customer Service or a technician authorized by
us (applies only to commercial users but is compulsory to comply
with the German Medical Products Act).
8.1.
L
R
AMP
EPLACEMENT
To ensure a high level of radiation intensity at all times we recommend replacing the
lamps after approx. 300 operation hours.
Disconnect the appliance by unplugging the power-pack!
Use a suitable screwdriver to undo and remove the three screws from the housing,
place the filter glass on a soft clean cloth where it cannot be scratched, and lift off the
top half of the housing. With a turning action you can now take the lamps out of their
holders and replace them with new ones. Re-close the housing, put the filter glass
back in the opening, and screw the housing tight.
Safe operation of the dermalight
filter is fitted!
Only use original type UV-B-311 nm (Order No. 19174) as spare parts!
I
NFORMATION
, R
EPAIRS
®
80 is guaranteed only when an undamaged
11