1. The main body of the container to our product is made of high—quality stainless
steel with one portion of Gr, eighteen of Ni, nine of Ti. Our product is not only
acid—proof , alkali--proof,but also safe and
three times longer than aluminum metal alloy product.
2. Our product heated up by immersion electric heating tube, consumes little power,
while temperature goes up quickly.
3. Vary heating method available, such as liquefied gas, coal gas, coal gas
stove ,firewood,etc.
4. The safety valve can release over-high pressure automatically when the steam
pressure exceeds 0.165MPa to insure the safety.
5. The pressure gauge is designed to have double scales, with two types of reading to
show pressure and temperature clearly.
1. Capacity of disinfect cylinder: 0.018 m
2. Working pressure rating: 0.14MPa
3. Maximal safety pressure: 0.165MPa
4. Power source pressure: 220V
5. Power of electric heating tube:2.0kW
1.Take out the product from the packing box properly.
2.The apparatus must be installed on flat surface; do not fix it use on unleveled
surface.
3.The left, right and back sides of the apparatus should be kept away from walls by
15cm at least.
4.To be sure the power source voltage is in accordance with the demand of the
apparatus, and dependable earth wire required.
5. Don't locate the apparatus near inflammable gas sources; it should be installed in
dry and ventilate environment.
Structure characteristics

Technique parameters

Installation and debugging

durable to consume, its life is
0.024 m
0.030 m
3
3
- -
3
3