Drawell DW-TDZ5-WS Manuali operativi - Pagina 4

Sfoglia online o scarica il pdf Manuali operativi per Apparecchiature di laboratorio Drawell DW-TDZ5-WS. Drawell DW-TDZ5-WS 20. Benchtop low-speed centrifuge

Drawell DW-TDZ5-WS Manuali operativi
model when replacing it.
* The centrifuge is not designed for the flammable and explosive matters. You
cannot make centrifugal for the matter (chloroform, acetaldehyde), and cannot
put the matters in the centrifuge or store them around centrifuge within 30cm.
Attention for the safe operation
* Please use the rotor and accessories which are designed for this centrifuge.
* Please check rotor, bucket and other accessories carefully before operation.
* Input the right rotor number before operation for not exceeding of its max
speed.
Attention: The max speed of rotor, buckets and accessories may be
different.
* Make sure that rotor can be installed on main spindle correctly for mot
damaging the spindle part.
* Put all buckets (the same model) in rotor before operation for not exceeding
its max imbalance volume, close the rotor lid then.
* Make sure is the centrifugal chamber is cleaned before operation.
* Don't reduce the speed by hand when it is rotating.
* Don't open door lid when rotor is rotating
* Don't move centrifuge lid when rotor is rotating.
* Anything is not allowed to enter into centrifuge during operation.
* If tube broken in centrifugal chamber, please clean out the fragments before
next operation.
* If there is water condensing in centrifugal chamber, please wipe up it with
clean cloth.
* The container with liquid should not be put around centrifuge; the electric
parts may be damaged for water spraying in.
* If centrifuge is irradiated by ultraviolet radiation for long time, the outer skin
may be faded. Clothing should be covered for avoiding irradiation.
* This centrifuge must not be used to separate the sample which producing
flammable and explosive gas. The stock around centrifuge is also forbidden.
* It is forbidden to operate the rotor exceed of its max speed. If user has not
set rotor type according with the number on rotor lid, all damages caused by
2