Warning: The ground wire should be connected properly in case of electric
shock.
As for measures, this centrifuge is adopted with single phase and three
wires plug to make sure of grounding. If your plug can not plug in the mains
jack, please contact the electrician to replay your main jack. Make sure that the
centrifuge is grounded properly.
5.2 Adjusting Requirement
Checking the goods:
Check the packing list the and make sure if damage is occurred during
delivery.
Installation , adjustment and maintenance should all be guided or operated by
professional technicians.
Door lid opening:
After installing the main machine, check the connection of all connectors and
fuses first, and turn on the main switch then. When the screen display
model/Logo, and then the main specifications will be showed. Pull the latch on
the door lid lock device by manually, and the door lid gradually opens at the
holding of gas springs polar and your hands,
Rotor installation:
Rotor should be checked carefully before using. It is forbidden to use rotor
and buckets with crack and corrosion spots (especially the buckets bottom of
swing rotor). All accessories should not be used if it is exceeding of its service
life. The conical surface of driving shaft should be smeared with some
lubricating grease if the main spindle is leaving unused.
Warning: The conical surface of rotor and driving shaft must be connected
closely.
As shown in the following figure: Firstly clasp the symmetric spoke of rotor,
7