Drawell DW-L Manuale operativo - Pagina 3

Sfoglia online o scarica il pdf Manuale operativo per Bilance Drawell DW-L. Drawell DW-L 6.

5.2. The Max weighing is the showing largest value after turn on add 9d
(e=10d,"d" is minimum reading). If overloading,it will show "------"and you
must take away the goods at once, or the balance (load cell,especially) will
be broken.
6. Tare
Press
, the balance pan can be tared if the blank spot disappears.
But if there is the black spot, the tare function is no effect.
7. Backlight
Press
to turn on/off the backlight.
8. Zero-Tracking and Auto Tare
8.1. Turn off the balance, one hand press and hold
another hand turn on the balance again, "-Zero-" will show, then loose

11. Maintenance

No light
Upper "----"
only
S-CAL
Value
unstable
Value
inaccuracy
or big error
at the same time,
Stay or show
wrong
symbols
Show "·" on
the
long
repeatedly
Show
Cou"
No data
No back to
zero
No
calibration
No Backlight
Failure
Cause
·Scale is not connected with
power
·Power switch is off
·Adapter is broken
·PCB is broken
·over-load
·load cell wire loose
·load cell wire loose
·No connect load cell wire
·Big air
·Working stage is unstable
·Big
change
of
temperature
·Load cell is broken
·No clear zero before weight
·No calibrate or calibration
weight is inaccuracy
·Wrong voltage
·Big error in four corners of
load cell
·Load cell is broken
·Short distracting outside
·Bad operation environment
left
all
(such as big air,vibration,big
or
change in room temperature
and so on.)
·No
set
constant
"No
counting
·constant is too big or small
·Distracted (by static... ) and
crash
·Wires loose
·PCB is broken
·Load cell is broken
·Wires loose
·Backlight is broken
Remedy
·To plug in the power
·To press
·To change new adapter
·To change new PCB
·To decrease load
·To re-connect the wire
·To connect load cell wire
·To shut door or window
·To put scale on stable
room
stage
·To control temperature
·To check load cell
·To press
·To re-calibrate
·To use right voltage
·To check error of four
corners
·To check load cell
·To reboot or re-plug in
power
·To
reboot
after
30
minutes
·To
control
working
temperature
before
·To set average
value
constant
·To re-calibrate
·To reboot or re-plug in
power
·To check all wires
·To change PCB
·To check load cell
·To check all wires
·To change new backlight