Drawell DW-LBI-150 Istruzioni - Pagina 6
Sfoglia online o scarica il pdf Istruzioni per Accessories Drawell DW-LBI-150. Drawell DW-LBI-150 11. Biochemical incubator
Anche per Drawell DW-LBI-150: Manuale d'uso (9 pagine)
disinfection.
E: Illumination key: Press to start or stop the illumination.
Parameter specifications:
1. In the standard state, long press the SET button, the upper row on screen displays LK,
adjust LK=8, press the SET button to enter setting
Symbol
Item
Power
KS
on/Start
Program
md
mode
Timing
tn
method
Hn
Timing Unit
High
and
Fn
low
fan
switch
Lighting
Ly
delay
Disinfection
srt
delay
Scheduled
boot
time
setting/
mon
scheduled
boot month
setting
2. Temperature parameter setting: In the standard state, long press the SET button, the
upper row on screen displays LK, adjust LK=3, press the SET button to enter the
temperature setting.
Value
0: Stop state when power is on 1: Run from
the first stage when power is on; 2: Run from
0~2
the place where the power was last cut off
after power on.
0: Customized control; 1: Multi-segment
0~1
programming control.
0:Start running timer, 1:when temperature
0~1
stable start timing (only fixed value)
0~1
0:minutes,1:hour
Start
the
temperature is between the( set value + Fn)
0~100
and (the set value-Fn); start the high-speed
fan when the temperature is outside (the set
value + Fn) to (the set value-Fn)
The lighting lamp is turned on,it will
automatically turn off when the delay time is
0~9999
not 0. when "Delay time is 0 ",the delay is
invalid, the lighting must be turned off
manually.
The disinfection lamp is turned on, when
the delay time is not 0, it will automatically
0~9999
turn off when the time is up. When delay time
is 0 , the delay is invalid, the lighting must be
turned off manually.
Function one; the meter without Beijing time
will automatically turn on after mon minutes;
Function two; the instrument with Beijing
0~9999
time, when Beijing time reaches the set time
(day, Hr, mTn) here, it will be turned on,
when Mon is greater than 12 ,the same as
function one.
4
Description
low-speed
fan
when
Factory set
value
0
0
1
0
the
1
0~9999M/h
(0)
0~9999M/h
(0)
0