Drawell SMART Series Manuale d'uso - Pagina 10

Sfoglia online o scarica il pdf Manuale d'uso per Microscope Drawell SMART Series. Drawell SMART Series 12. Biological microscope

SMART Series Biological Microscope
Trouble
One side of image is
dark or image is
moving as focusing
Objective touches
specimen as changing
low times objective to
high times objective
Image observed by two
eyes aren't in
superposition entirely.
It is easy for eyes to be
tired during observing
8. SMART Series Biological Microscope Outfits
Items
Eyepiece
Finity Achromatic
Objective
Finity Plan
Objective
Infinity E-Plan
Objectives
Infinity Plan
Objectives
Seidentopf Binocular Head
Seidentopf
Trinocular Head
Nosepiece
Mechanical Stage
Causation
There is bubble in immersion oil
Size of iris aperture isn't
proper
Position of condenser is too low
Objective
isn't
position
Specimen isn't placed correctly
Cover glass of specimen isn't in
up direction
Cover glass doesn't meet the
requirement
Interpupilary
distance
adjusted correctly
Diopter isn't adjusted correctly
Specification
WF 10X-18mm
WF 10X- 20mm
4X
10X
40X(S)
100X/1.25 (Oil) (S)
4X
10X
40X/0.65 (S)
100X/1.25 (Oil) (S)
4X
10X
40X (S)
100X (Oil) (S)
4X
10X
40X (S)
100X (Oil) (S)
Plan 20X
Plan 60X (S)
Inclined 30°,Rotatable 360°,
Interpupilary Distance 48-75mm
Inclined 30°,Rotatable 360°,
Interpupilary Distance 48-75mm,
Light Distribution 20:80
Quadplex , Quintuple (Optional)
Stage Size: 145X140mm
Travel: 76mmX52mm, Scale: 0.1mm
Coaxial Coarse and Fine
Focusing Knobs
Travel Range: 26mm, Scale: 2um
Two Slide Holder
Integrated stage, Stage size:
182X140mm
-9-
Clear the bubble way
Adjust the size of iris
aperture
Readjust the position of
condenser
in
correct
Make the objective in
correct position
Place specimen levelly on stage
and clip it with clamp
Place specimen correctly
Use required thickness
cover glass (0.17mm)
isn't
Adjust interpupilary
distance according to two eyes
Readjust diopter
SMART-1 SMART-2 SMART-3 SMART-4
●●
Remedy
●●
●●
●●