4. Lift the logic board and inspect for damaged or separated bezel posts.
If there are any damaged or separated bezel posts, the bezel must be replaced. If there
are no damaged or separated bezel posts, continue to Step 5.
Damaged
bezel post
PN P00000360
Record number 1103-001-008-R-SD-INF, DIR 10000307230, Version 01 Release Date: 16AUG17
DOCO Use Only (Approver's List): 0207-001-001-S (Latest Version) to determine authorized approvers
DOCO Use Only (Record Format Template) 0204-011-005-R, DIR 10000153640, Ver 02, Release Date: 04APR16
Inspecting the Alaris™ Pump Module Bezel
DIR No.: 10000351665 Version: 00
Bezel posts
Separated
bezel posts