8. Close the hood and ensure it snaps into place. This will allow the laser interlock to
disengage and allow laser light to pass through the flow cell.
Setting up the break off point
1. . Make sure that the stream appears in the Break off Window.
a. No stream, dripping or spraying: remove nozzle tip. Sonicate the nozzle,
wipe out the flow cell and sort chamber.
b. Unsteady stream: shut off the stream, turn it on again. If problem persist,
remove the nozzle tip, sonicate, and put it back on again.
c. Stream obscured by grey area: sheath fluid in front of droplet camera. Turn
off the stream, remove the nozzle tip and the key, dry everything out, and
clean the chamber with kimwipe.1
2. Monitor the droplet formation
Gap
Drop 1
Satellite
droplets
70um tip
85/100/135um tip
3. The leftmost figure above shows fast converging droplets wherein the satellites
attach to the top of their droplet. Fast converging droplets are required to use the
70um tip and are preferred for the 100um tip. Slow converging droplets, shown in
the rightmost figure, are never preferred but can be used for the 100um tip. If you
observe slow converging droplets while using the 70um, remove and sonicate the tip
as slow converging droplets can never be used with the 70um tip.
Gap (wide band): distance between the
unbroken stream and the first unattached
droplet
Drop1 (thin band): position of the first
unattached droplet
70/85um tip: satellites attach to the droplet
on top, the gap is located at between the
above satellite droplet and the following
droplet.
85/100/130um tip: Satellites attach the
droplet below, gap is located just on top of the
last attach droplet and its following satellite
droplet