Alentec & Orion 24702 Manuale d'uso - Pagina 6
Sfoglia online o scarica il pdf Manuale d'uso per Strumenti di misura Alentec & Orion 24702. Alentec & Orion 24702 8. Hose end meter pre set version
Manual 24702_EN_2B
Instructions for re-calibration
1. Check the Meter against a 2-liters measure.
If the reading on the Meter is out of calibration, note the
displayed volume (e.g. 1.98) and calculate the new Scale
Factor (e.g. 1.0101) as shown in illustration to the left.
Note! If the meter already has a Scale Factor other than
the default figure 1.0000 set it back to 1.0000 (see point
3-5.) before performing the calibration. To check the
actual Scale Factor, simultaneously press the RESET
1 0 1 0 1
and SELECT buttons (press RESET slightly before
1 9 8
SELECT).
New Scale Factor
2. Remove the hexagon screw that covers the SETUP
button.
3. With a sharp object, like a small screwdriver, short-
circuit the two soldered pads on the circuit board that
can be reached through the hole. These must remain
short-circuited during the entire setting procedure.
4. The actual Scale Factor is now shown on the display.
Set the new Scale Factor as shown in the diagram. Use
the SELECT button to move to desired position and the
RESET button to change the numbers.
5. Release the SETUP button and test the Meter against
the 2-liters measure. If the Meter reading is accurate,
replace the hexagon screw.
6(8)