Sfoglia online o scarica il pdf Manuale O & M per Trasmettitore ATI Technologies B12-69. ATI Technologies B12-69 20. 2-wire wet oxidant gas transmitter

Model B12-69 Wet Oxidant Gas Transmitter
The only maintenance required on the B12-69 system is periodic sensor service. The wet
oxidant gas sensor uses a microporous membrane that can lose sensitivity with the accumulation of
solids or precipitates in the membrane. As a preventive measure, sensors should be rebuilt with new
membrane and electrolyte about every 3-4 months.
Follow the sensor assembly procedure on pages 10 and 11 of this manual to replace the
membrane and electrolyte.