Model B12-70 Wet Acid Gas Transmitter
The only maintenance required on the B12-70 system is periodic sensor service. The wet acid
gas sensor uses a microporous membrane that can lose sensitivity with the accumulation of solids or
precipitates in the membrane. As a preventive measure, sensors should be rebuilt with new membrane
and electrolyte about every 3-4 months.
Follow the sensor assembly procedure on pages 10 and 11 of this manual to replace the
membrane and electrolyte.
O & M Manual
-Release E, 7/15
MAINTENANCE
- 17 -