10. TROUBLESHOOTING (1)

PROBLEM
LCD Display shows
battery symbol
LCD Display shows
"EE"

11. TROUBLESHOOTING (2)

PROBLEM
LCD Display shows
"EE"
LCD Display shows
abnormal result
No response when you
press button or load
battery
POSSIBLE CAUSE
Low Battery
Arm or blood pressure
monitor was moved
during testing
The cuff does not inflate
properly or pressure falls
quickly during testing
Irregular heartbeat
(arrhythmia)
POSSIBLE CAUSE
The cuff was not properly
applied or the rubber
tube was bent or
pressed.
The cuff position was not
correct or it was not
properly tightened
Body posture was not
correct during testing
Speaking, arm or body
movement, angry,
excited or nervous during
testing
Incorrect operation, or
strong electromagnetic
interference
10/15
SOLUTION
Change all the batteries
Re-test taking care to not
move your arm or the
blood pressure monitor
Make certain the rubber
tube is fully inserted into
the blood pressure
monitor
It is inappropriate for
people with serious
arrhythmia to use this
blood pressure monitor.
SOLUTION
Review the cuff applying
and testing sections of
the instructions and
re-test.
Apply the cuff correctly
and try again
Review the body posture
and testing sections of
the instructions and
re-test.
Re-test when calm and
without speaking or
moving during the test
Take out batteries for five
minutes, and then
reinstall all batteries.
Template version:1
Operation Guide