NOTE: Once you select the start button an audible exit alarm will begin to sound and the Sanidyne
will begin counting down from 1:30 seconds to allow all inhabitants to leave the area before the UV treatment begins.
The words "LEAVE AREA" will now appear on the lower status bar of the Hand-held Remote Controller. (See Figure
3, for the specific location). The Occupancy Sensors located on the top of the unit (Shown on Figure 1) will begin to
operate at this time, providing an additional safety area around the unit. If the sensors detect movement close to the
Sanidyne
Prime Remote when less than 30 seconds remains on the countdown time, the Sanidyne
®
will automatically reset giving 30 seconds more to leave the Operation Area. The Countdown time as well as the Treatment
Time will all be displayed on the Hand-held Remote Controller. See Figure 3, for the specific location.
Occupancy Sensors do not protect the entire Operation Area of the Sanidyne
circumstances should this unit be permitted to operate with human, or animals present in the Operation Area.
4.
The Hand-held Remote Controller displays the Lamp Status; When the lamp operation indicators on the display glow yellow
the lamp is on, when the lamp is off it displays grey. (See Figure 3 to locate lamp operation indicators) Once the treatment
time has expired the unit will turn off the Germicidal Ultraviolet Lamps and it is now safe to return to the area.
If at any time during the Treatment Time the Occupancy Sensors detect someone near the
Sanidyne
Prime Remote
®
the
Sanidyne
Prime Remote
®
display or at the
Sanidyne
Sanidyne
Prime Remote
®
again to begin the 1:30 countdown to allow all inhabitants to leave the area. After the 1:30 has elapsed the unit will power
on and complete the remaining Treatment Time left prior to the alarm.
If the Hand-held remote is used to stop the alarm, the operator must return to the area and ensure that the area is safe to
proceed. Select the START/STOP button on the keypad controller to begin the 1:30 countdown to allow all inhabitants
to leave the are. After the 1:30 has elapsed the
Treatment Time left prior the alarm.
s
D
etting
ate anD
Press the Settings Button.
1.
Press Set Date/Time.
2.
Press Admin Pin. Using keyboard enter Pin number 1234.
3.
Press Date or Time.
4.
5.
Scroll in Date or Time by swiping the scroll bar.
6.
Press Done. Then press Home button when finished.
s
s
aving a
ession
Saving a Session: At the end of a Session or after a Session is Stopped the "Save This Session" box will appear.
To save a Session;
Press Yes under Save Session.
1.
Press on Note Entry
2.
Type in Session Notes, press OK
3.
Clearing Total Op Time or Clearing Session Data:
1.
Press Settings Button
2.
Press Settings
3.
Press on the box next to "Reset Total Operating Time" and /or "Clear Session Data."
4.
Press Home Button
12
, the unit will immediately turn off and send an alarm to the Hand-held Remote Controller and
. If this event occurs the alarm can be turned off on the remote by pressing anywhere on the
by pressing the START/STOP on the Keypad. If the Alarm is turned off at the
Prime Remote
®
the operator can reset the unit by pressing the START/STOP button on the keypad controller
Sanidyne
t
ime
will power on and complete the remaining
Prime Remote
®
Prime Remote
®
Prime Remote
®
Prime Remote. Under no
®