PRINCIPLE OF OPERATION
Best practices when checking a lamp should include the following data:
•
Lamp Description or Part Number
•
Displayed value indicated on the
•
Distance at which measurement was taken (Ref: 1 meter)
•
Lamp warm-up time used before lamp was tested
*Note: Your lamp values may vary from lamp manufacturer's data since Lamp Manufacturer's output data is based on
measurements performed under laboratory conditions in air at ambient room temperatures, utilizing matched electronic ballast
and represent average values.
Equipment is required to be powered on and in use when testing for UV-C Germicidal Leakage.
It is the user's responsibility to ensure inspection in and around these fixtures is safe to perform.
When using this device to check a UV-C source, avoid exposure to direct or reflected germicidal
ultraviolet rays. Germicidal ultraviolet rays are harmful to the eyes and skin. Wear face, hand, and eye
protection and cover any skin that may be exposed to direct or reflected germicidal ultraviolet rays.
M
uV-C G
easurinG
Fixtures are often inspected to ensure the UV-C light doesn't exit from within framework of the device, where it could harm
occupants or damage other susceptible material located in the vicinity. This type of test differs from the testing of an individual
UV-C Germicidal lamp in that the distance and angle from the source plays a much smaller role.
Using the
UV-C Detector, test along the fixture's seams, covers, duct work, light traps, or any area which appears that
Zenith
TM
light can be escaping. Ideally, the
may inherently not allow for complete blockage of UV-C light, such as those found over conveyor systems or tunnels. For these
types of fixtures, it may be suitable to develop a safe zone boundary around the fixture. Contact one of our Atlantic Ultraviolet UV
Application Specialists at 1-631-273-0500, Monday–Friday 8:30am to 5pm EST for further information on this topic.
6
UV-C Detector's digital display
Zenith
TM
F
l
erMiCidal
ixture
Zenith
UV-C Detector should display "0" in all cases and at all distances. Some fixture designs
TM
(Cont.)
eakaGe