6. PERIODIC INSPECTION

The device is designed to operate for 2 years under normal operating conditions, based on a daily
usage rate of 8 infusions per day, i.e. more than 6,000 times. However, the manufacturer advises that
users perform periodic maintenance and inspections to ensure the correct operation of the device,
either by sending it to the manufacturer or distributor, or by using the method described below:
Place a 1000 ml Viaflo infusion bag of NaCl 0.9% marketed by the trademark Baxter in the Droper Field
and introduce the piercing pin of an infusion line without a catheter in the classic OFF position. Drain
the bag and line as described in 2.4 Draining the infusion line and bag. Return the Droper Field 1000
to its horizontal position. Open the infusion line into a recipient placed at the same height as the device
and measure the time required to drain the bag. This time must be between ±320 and ±340 seconds.

7. STORAGE

The device must be stored in a temperature-controlled, dry place.
Recommended relative humidity: 20% to 90%
Recommended storage temperature: - 10°C to + 60°C

8. WARRANTY CONDITIONS

The Droper Field 1000 is guaranteed against any part or manufacture defects for a period of 1 year
from its invoice date. To benefit from the parts and labour warranty from our After-Sales Service or
from an approved service, the following conditions must be complied with:
The device must have been used under the normal operating conditions described
herein.
The device must not have suffered from any deteriorations caused by its storage,
maintenance or incorrect handling.
The device must not have been modified or repaired by persons not authorised by
PROMOVET.
The serial number of the device must not have been modified or erased.
For item returns and repairs, please contact the After-Sales Service.