Motor Assembly Replacement
A2000/A2000I
To replace the Motor in the Motor Assembly-
1. Loosen the two motor pulley set screws.
2. Loosen the four mounting bolts for the motor. Slide the
motor
forward and remove the drive belt and motor
pulley.
3. Remove the motor mount bolts. Replace the motor oriented
as below. Install the pulley and install the motor mount bolts-
leave them loose enough to slide the motor and to tension the
drive belt to a 1/8" (3mm) deflection. Tighten the mount
bolts.
4. Install the motor assembly, tighten the mounting nuts and
coupler set screw.
5. Re-connect the wiring.
6. Replace the back and bottom panel and the instrument will be
ready for use.
023
Service Procedure