Ankom Daisy D200 Sostituzione
Sfoglia online o scarica il pdf Sostituzione per Accessories Ankom Daisy D200. Ankom Daisy D200 4. Motor replacement
To replace the Motor on the ANKOM Daisy
Note: The following items will be sent in a replacement package: New Motor (120v or 220v), fan for the 120v motor.
1.
Prepare the instrument for service.
a. Power off the instrument.
b. Remove the back panel of the Daisy cabinet using a Phillips
screwdriver. Set aside the four screws for later use.
2.
220v Motor: Disconnect the two motor wires from the Terminal
Block Assembly.
*Refer to the labeled image (right) of a Terminal Block Assembly
for the following steps in this procedure.
a. Insert a small flat-blade screwdriver into the correct terminal
blocks to unscrew the screw inside the circular holes and
loosen the wires.
b. Unscrew the white wire from the #1 gray terminal block and
the red wire from the #3 blue terminal block.
Note: Refer to the manual for further wiring instructions.
3.
120v Motor: Disconnect the four motor wires from the Terminal
Block Assembly.
a. Insert a small flat-blade screwdriver into the correct terminal
blocks to unscrew the screw inside the circular holes and
loosen the wires.
b. Unscrew the blue and white wires from the #1 gray terminal
block and the brown & black wires from the #3 blue terminal.
Note: Refer to the manual for further wiring instructions.
ANKOM Technology 2052 O'Neil Road Email: [email protected] Phone: 315-986-8090 Fax: 315-986-8091
Motor Replacement
D200/D200I
II
Incubator, follow the steps below.
Service Procedure
Revised: 08/06/2014
Terminal Block Assembly
Blocks 1-6
1
2
3
4
Terminal #1
Terminal #3
Terminal #1
Terminal #3
001
5 6
pg. 1 of 4