3M 718A Manuale d'uso - Pagina 10

Sfoglia online o scarica il pdf Manuale d'uso per Strumenti di misura 3M 718A. 3M 718A 14. Static sensor
Anche per 3M 718A: Manuale di istruzioni (12 pagine)

3M 718A Manuale d'uso
5.2
Calibration: the 3M
718A products are supplied by the factory pre-calibrated. 3M does not
specify a minimum calibration cycle for the Static Sensor 718 or Air
Ionizer Test Kit 718A products. The user, usually according to internal
Quality procedures, determines calibration cycles. In the event that the
user wishes to perform a self-calibration, the following steps should be
followed for the Static Sensor 718 (user-calibration not possible on the
Air Ionizer Test Kit 718A).
Equipment Needed:
5.3
Test Fixture High-Voltage Power Supply, capable of supplying voltages
up to 10,000V
Voltmeter, with > 50 kΩ input impedance, capable of measuring voltages
down to the μV range.
Cable with a 3/32 inch (2.5 mm) mono plug and secondary connector to
interface with voltmeter.
5.4
Test fixture:
Metal plate of at least 5 square inches area
(38.7 cm
) area.
2
Metal stand capable of supporting Static Sensor
718, and holding it one inch (2.54 cm) away
from the metal plate, centered.
Connectors on the plate with which it can
interface with the high voltage power supply.
Connections on the metal stand with which it
can be connected to electrical ground.
5.5
Procedure:
5.5.1 Place the Static Sensor Static Sensor 718 on the
metal stand. Verify that it is exactly one inch
(2.54 cm) away from the metal plate, and that its position is centered
relative to the plate.
5.5.2 Connect the stand to ground.
5.5.3 Turn on the Static Sensor 718 and set it to the low range.
5.5.4 Ground the plate. Zero the display.
5.5.5 Remove the ground from the plate, and connect it to the high voltage
power supply. Apply a 1 kV charge to the plate.
Static Sensor 718 and 3M
9
Air Ionizer Test Kit