3M Clean-Trace Luminometer LX25 Manuale d'uso - Pagina 14

Sfoglia online o scarica il pdf Manuale d'uso per Strumenti di misura 3M Clean-Trace Luminometer LX25. 3M Clean-Trace Luminometer LX25 16.

3M Clean-Trace Luminometer LX25 Manuale d'uso

Troubleshooting

This section is intended to help solve problems that may be encountered when using the Clean-Trace Luminometer LX25 .
Contact 3M or your local 3M distributor for more information .
Problem
Diagnostic Check Failed
Battery Low
Chamber Door Open
No Swab Detected
Instrument temperature is not
within normal operating range
Swab Left in Chamber
No Display
Results Lower than Expected
Results Higher than Expected
Other Errors
Probable Cause
Chamber Door is open
Operating Clean-Trace
Luminometer LX25 out of
correct temperature range
ATP Test UXC or H2O left in
ATP test chamber
Battery needs to be charged
Chamber door was open
while attempting to start the
reading process
Instrument temperature is
>40°C or <5°C
Battery fully discharged
Clean-Trace
Luminometer LX25 is in Sleep
Mode or has been turned off .
Instrument malfunction
Sample / reagent problem
Instrument malfunction
Direct sunlight or static
discharge can cause ATP Test
to phosphoresce and cause
high readings
Remedy
Close the chamber door and retry the Diagnostic Check
Operate the Clean-Trace Luminometer LX25 between 5°C and
40°C . Retry the Diagnostic Check .
Remove the ATP Test UXC or H2O from the test chamber and retry
the Diagnostic Check .
If the Diagnostic Check fails after a retry, contact 3M or your
local distributor .
Connect the power cord cable to the Clean-Trace
Luminometer LX25 and connect the cord to a power outlet to
recharge the battery .
Close the chamber door after inserting a Clean-Trace ATP Test UXC
or H2O and before starting the reading process .
Insert a Clean-Trace ATP Test UXC or H2O .
Use the Clean-Trace Luminometer LX25 at ambient temperature
between 5°C and 40°C . If the problem persists, contact 3M or your
local distributor .
The Clean-Trace Luminometer LX25 has a sensor to determine if a
Clean-Trace ATP Test UXC or H2O has been left in the chamber for
more than 30 seconds following the completion of a measurement .
If this situation is detected, the instrument will display a "Swab
Left In Chamber" error message on the screen and provide an
audible beep to remind the user to remove the Test following
the measurement .
Recharge the battery .
Press the "Up" button on the cursor control to wake from sleep
mode . If no response, press the "Power" button for 3 seconds to
turn on the Clean-Trace Luminometer LX25 .
Contact 3M or your local distributor .
Check to ensure correct sample technique was used and ATP Test
UXC or H2O activation was followed according to Clean-Trace ATP
Test instructions for use . Verify with known positive samples such
as a 3M Positive Control Test .
Contact 3M or your local distributor .
After exposure to sunlight or static discharge an immediate repeat
reading is 50% of the first reading, if this is the cause of the higher
than expected results . Subsequent readings become incrementally
lower and finally return to normal .
Avoid exposing ATP Test UXC or H2O to direct sunlight or
static discharge .
1 . Turn off the Clean-Trace Luminometer LX25 and power back
on again .
2 . Retry ATP Test UXC or H2O using the Clean-Trace
Luminometer LX25 .
3 . If the problem persists, contact 3M or your local distributor .
14