Drucker Diagnostics 614VET Manuale dell'operatore - Pagina 5

Sfoglia online o scarica il pdf Manuale dell'operatore per Apparecchiature di laboratorio Drucker Diagnostics 614VET. Drucker Diagnostics 614VET 16.

Drucker Diagnostics 614VET Manuale dell'operatore
G
S
ENERAL
PECIFICATIONS
Nominal Speed (Fixed Angle)
Nominal RCF (Fixed Angle)
(RCF calculated at a rotor radius of 4.23 in. with 7713032 tube holders)
Maximum capacity (Fixed Angle)
Overall Dimensions (H x W x D)
Centrifuge Motor:
Maximum Acceleration Time
Protection Breaker
Timer
Current Requirement
Voltage Requirement
Frequency
Weight
Any use other than those specified by the Manufacturer is explicitly prohibited.
Maximum sample density is 1.15 grams / mL (water density = 1.0 grams / mL)
S
L
ETUP
OCATION
1. Unpack the centrifuge and verify that all supplied equipment is
present.
2. Choose a setup location which meets the following criteria:
a.
A clearance height of 20" is required to open the lid.
b. The clearance envelope is the space around the centrifuge
which is required for safety. Choose a setup location which
will allow for a clearance envelope of at least 24" x 24",
(with the centrifuge at the center). No person or hazardous
material shall be permitted in the clearance envelope during
operation. The operator time within the envelope shall be
limited to the time necessary for loading, unloading and
centrifuge operation only.
c.
Proper ventilation is necessary to prevent the overheating of
samples as well as premature failure of the centrifuge.
Choose an area which will allow unencumbered air flow.
d. The centrifuge is designed to secure to the operating surface
by four suction feet. No adjustment is necessary for leveling
the centrifuge; however, the surface should be flat and level.
e.
Be sure the outlet is always within reach as the line cord is
the means of emergency disconnection!
Page | 5 Drucker Diagnostics +1-814-692-7761 - [email protected]
1,750 (+ 50 / - 10) RPM
370 xg
90 mL (6 x 15 mL)
8.75 in. x 11.75 in. x 14 in.
Brushless P.S.C. motor
10 seconds
5 Amp. re–settable
Electronic, 1 to 30 minutes
preset to 2 minutes, +/– 2%
1.9 Amps
115 (+/- 10) Volts
60 Hz
11 lbs