5 TROUBLESHOOTING
PROBLEM
The lid does not open.
PROBLEM
Excessive vibration
PROBLEM
Rotor does not spin
PROBLEM
Clicking noise during braking
PROBLEM
Whistling noise while running
DRUCKER DIAGNOSTICS
MODEL 642E SERVICE MANUAL
POSSIBLE CAUSE
No Power
No Power
No Power
Lid knob is ajar
Lid lock is active (Unlock timed out)
Lid tray is unplugged from PCB or defective
PCB is damaged
POSSIBLE CAUSE
Rotor improperly loaded
Debris lodged within the rotor or tube
carriers
Centrifuge housing is loose
Missing/damaged feet
Motor failure
Rotor damaged
POSSIBLE CAUSE
No Power
No Power
No Power
Lid not properly latched
Internal connection failure
PCB failure
Motor Failure
POSSIBLE CAUSE
Rotor is loose
POSSIBLE CAUSE
Debris in air intake / exhaust ports
Gasket failure
Gasket failure
SOLUTION
Check line cord
Check circuit breaker on underside of centrifuge.
Check wall outlet
Rotate the lid knob fully clockwise before pressing the
'OPEN' button
Press the 'OPEN' button to de-activate the lid
Requires service
Requires service
To gain access to the rotor - Remove the 'OPEN/CLOSE'
sticker and slide the lid latch lever toward the front of the
centrifuge. This will unlock the lid.
SOLUTION
Load equally filled tubes symmetrically in the rotor. All
carriers and/or tube holders must be present in the rotor,
whether loaded, or empty.
Carefully inspect all rotor pockets, tube holders and
crevasses for debris.
Requires service
Requires service
Requires service
Replacement required
SOLUTION
Check line cord
Check circuit breaker on underside of centrifuge.
Check wall outlet
Press down firmly on lid and rotate lid knob clockwise until
the 'LATCHED' light illuminates.
Requires service
Requires service
Requires service
SOLUTION
Tighten rotor nut per section 7-2
SOLUTION
Remove power before clearing debris.
Requires service
Requires service
SM006
REV: A