significant difference in zinc-air battery life, especially in hot and humid weather. Individual results will vary. Here are
specific suggestions:
•
Behind-the-ear speech processors: Leave zinc-air batteries in the processor (where applicable) but remove
the battery cover. Place the processor, headpiece and cable in the drying compartment on the right. Do not put
anything except the desiccant on the left side of the tray.
•
Body-worn speech processors: Place as many components in the drying compartment as will fit without
crowding. Size limitations might mean that the processor will need to be conditioned separately from other
hardware. If this is the case, the microphone assembly should be conditioned more frequently than the processor,
as it is more susceptible to moisture damage. The battery pack may need to be removed from some processors.
General product care
To keep your Dry & Store working properly, do not operate it in high-humidity locations such as a bathroom. Keep the
lid closed at all times, because the effectiveness of the desiccant will be reduced due to moisture adsorption from the
surrounding air. Clean with a soft cloth. Never use strong cleaning agents or abrasives. Do not spray any liquid into the
tray or onto the UV lamp.
troubleshooting
•
No lights illuminate after pressing the <on> button: Is the lid closed tightly and has the <on> button been pushed?
If yes, verify that the AC adapter is securely connected to both the Dry & Store Global unit and the outlet/mains
and that the outlet/mains is not controlled by a wall switch which is in the off position. Next, try another AC
adapter (available from the place of purchase, a service center, or at www.dryandstore.com). Replace it only with
an adapter provided by the manufacturer. If the unit still will not start, the internal electronics are probably faulty,
so return it to the place of purchase or service center for evaluation. NOTE: this condition does NOT indicate a
bad UV lamp.
•
No blue light is visible during first three minutes of the operating cycle, but the green light is on: Check that the
UV lamp is firmly positioned in its sockets. If not, remove and reposition the lamp, then re-start the unit with
the lid closed tightly. If the blue light does not illuminate during the first three minutes, replace the UV lamp
(available from the place of purchase, a service center, or at www.dryandstore.com). Replace it only with the
special lamp provided by the manufacturer. If both blue and green lights fail to light, see "No lights illuminate..."
above.
•
Unit is noisy: Noise problems are usually fan-related. Sound resonates inside the unit, so some noise is to be
expected, even from the high-quality ball-bearing fan used in Dry & Store. If the fan noise varies in intensity or
frequency, disconnect the unit from power and check for obstructions. If none, have the unit serviced.
•
Lights illuminate, but unit is not drying properly: Has the Dry-Brik been replaced regularly? It is important that
it be replaced every two months according to the instructions. Check for fan movement. If no fan sound can be
heard, return the unit to the place of purchase or service center for evaluation.
•
Metal bead inside UV lamp: This is perfectly normal with bulbs made by Sankyo-Denki or Ushio. It is a result of
their environmentally-sensitive manufacturing process.
In North America, call Customer Service at 1-888-327-1299 or visit www.dryandstore.com
Outside North America, contact the place of purchase or a service center for evaluation.
Rev B