3M SoundPro SE Series Manuale di avvio rapido - Pagina 3

Sfoglia online o scarica il pdf Manuale di avvio rapido per Strumenti di misura 3M SoundPro SE Series. 3M SoundPro SE Series 6. Sound level meters

3M SoundPro SE Series Manuale di avvio rapido
1.
Ensure the SoundPro is turned On and is either stopped or paused.
2.
Attach the calibrator and cal adapter to the SoundPro. Set the calibrator to 1 KHz and 114 dB (if it is a selectable).
Microphone
Place calibrator
over microphone
3.
From the Start screen, press the Cal softkey and the Cal screen will open. Press
screen will open.
4.
Allow the measurement to stabilize and press
Calibration history screen appears (see B).
5. Press
6. For a Post Calibration, ensure the SoundPro is in "Pause" mode. (Essentially, run a measurement and press the run/pause key.)
The Pause indicator will appear at the top of the display.) Then repeat the Calibration steps above and press Enter to store the Post
Calibration information. A sample screen is displayed below with post calibration history.
Position the SoundPro in the field. Press the run key to start measuring. To pause a study (for example: lunch breaks), press the run/pause key. To
start measuring again, press the run/pause key. Continue this process until your measurements are complete by press the Stop key. This will end
the session.
1. From the Start screen, select either View Session or select View Current Study by press the
2. Press
Lookup softkey:
Select to change
measurement
parameter (Lpk,
Lmx, Lmn,
Leq,Las)
Cal softkey
 Select softkey to open
A. Calibrate
Press corresponding
softkey to 114.0 dB
level.
key to exit the Cal screen.
key to begin measuring. Press
RunTime
Run icon
Measurement value (Average SPL)
QC10/QC20
Select 1kHz and 114 dB
Arrows to set 114.0 dB level (see A). Then, press
key to Pause (optional). At any time, press
when Calibrate is highlighted. The Pre-Cal
key. When set, the
B. Pre-Cal
Displays pre-calibration
information
arrows and
to end the session/study.
key to select.