3M Static Monitor 790 Manuale d'uso - Pagina 10
Sfoglia online o scarica il pdf Manuale d'uso per Strumenti di misura 3M Static Monitor 790. 3M Static Monitor 790 19.
a. Red lamp(s) flashing with an audible alarm.
This alarm mode can be caused by two conditions and both are interrelated. If
either the voltage level on the operator is greater than the pre-selected internal
reference voltage or poor contact between the wrist band and the arm exist, the
alarm will activate.
If the 3M
Static Monitor 790's alarm activates when the operator is not
™
moving (normally a non-charging situation), the cause is most likely due to
poor contact between the wrist band and the arm. The alarming period would
normally be long in duration (unless intermittent contact) during this type of
condition. Check contact between the wrist band and arm. Some operators
may have difficulty in providing sufficient contact to the wrist band because
of dry skin or arm hair. They may need to use an approved skin moisturizer or
reposition the wristband on the arm.
If the monitor's alarm does not activate while the operator is not moving, then
the cause of the alarm was most likely due to a voltage being developed on the
operator. A high resistance between the operator and ground would allow the
operator to charge up as they moved about. The alarm period for this type of
fault would normally be short and coincide with their body movement.
Alarming can also be caused by an intermittent or high resistance in the wrist
strap's dual conductor ground cord. Replace a suspect cord with a known good
ground cord or check the end to end resistance of the cord using an ohmmeter.
Approximately 1 megohm of resistance should exist between the tip to tip and
barrel to barrel contacts of the two plugs located on opposite ends of the cord.
Note: The operators may complain that the alarm is sounding too often until they learn to adjust the
wristband to fit securely or apply an approved skin moisturizer on a frequent basis. Please remember
that the monitor is informing you that the operator is exceeding the established static control
requirement for voltage to ground when wearing a static protective wrist strap assembly. These
alarms alert the operator when sensitive electronics are possibly being exposed to static electricity.
Prior to incorporating the static monitor into your static control process, the operator was unaware
of these events.
b. Two red lamps illuminated (continuously) with an audible alarm.
This indicates that a high resistance condition (greater than 10Ω) exists
between the Chassis/Ground Cord and ground. This will include the
equipment chassis ground if being monitored. Check the ground cords for
continuity and ground attachment points.
10