DS Maref DVT-2600 Manuale di servizio - Pagina 14
Sfoglia online o scarica il pdf Manuale di servizio per Apparecchiature mediche DS Maref DVT-2600. DS Maref DVT-2600 16.
11-4. Device error test
No
code
1
Sleeve connection
error
2
Lower Pressure error
3
High Pressure error
12. Maintenance Check list
No.
Classification
1
2
Controller
3
Visual Inspection
4
5
6
7
Labeling
8
Visual Inspection
9
Power cord
10
Visual Inspection
Tubing set
11
Visual Inspection
12
13
14
Accessories
15
Visual Inspection
16
17
18
19
20
21
Pressure Check
22
[LEG 40mmHg]
23
24
25
Pressure Check
Error
- Press start key for 40 seconds to check the condition of
wearing sleeves while the controller is not connecting to
sleeves. The error will occur if the controller doesn't
recognize the sleeves after 40 seconds.
- Remove the sleeve in the normal condition and check the
error after 3 cycles.
- While the controller is running normally, check the error in
the maximum 10 cycles after not allowing the air by
bending of tubing.
Item
Display
The control button
Air discharge
The bed hanger
Handle
Equipment label
Error sticker
Quick-guide
Battery charge-guide
Power cord
Tubing set
User manual
Quick guide
Sleeve manual
Thigh sleeve
Calf sleeve
Boot sleeve
Foot sleeve
1 Chamber - 1ch
1 Chamber - 2ch
1 Chamber - 3ch
2 Chamber - 1ch
2 Chamber - 2ch
2 Chamber - 3ch
1 Chamber - 1ch
test method
Inspection
Damage, Foreign Substances
Damage, Not pressed
Damage
Damage
Damage
Deleted, Blurred, Torn
Deleted, Blurred, Torn
Deleted, Blurred, Torn
Deleted, Blurred, Torn
Cut, Scraped, Pressed
Torn, Blocked, Damaged
Visual Inspection
Visual Inspection
Visual Inspection
Visual Inspection
Visual Inspection
Visual Inspection
Visual Inspection
45mmHg (39 ~ 51mmHg)
40mmHg (46 ~ 34mmHg)
35mmHg (41 ~ 29mmHg)
45mmHg (39 ~ 51mmHg)
40mmHg (46 ~ 34mmHg)
35mmHg (41 ~ 29mmHg)
120mmHg
Pass
Fail