10. Check the direction of flow:
Ensure the sensor elements are aligned correctly with the flow (±2 degrees). Note differences between
equal and unequal sensor types.
Insertion type flow meters: Is the arrow on the flow meter probe pointing in the direction of flow?
Inline type flow meters: An inline meter will have the flow direction arrow on both the probe and flow
body. Are both arrows pointing in the direction of flow?
11. What is the CSV voltage?
Press the F1 & F2 keys at the same time in normal mode to access the engineering displays. Display #10
will have the CSV value. Compare this value to the meter's calibration certificate.
F-184, Rev A