ATsens AT-Patch Manuale d'uso - Pagina 4
Sfoglia online o scarica il pdf Manuale d'uso per Prodotti per la cura della persona ATsens AT-Patch. ATsens AT-Patch 14.
Anche per ATsens AT-Patch: Manuale rapido (2 pagine), Manuale rapido (6 pagine), Manuale d'uso (14 pagine)
Attaching the sub patch
7
Put the sub patch on top of the attached device in the correct direction.
After attaching the sub patch, press firmly on the entire surface for at
least
1 minute
Removal of the protective tape
8
After attaching the sub patch, detach and remove the protective tape.
Checking the attachment of the device
9
Check if the device and the sub patch are attached well without any loose
parts.
6
02
Attach and Installation
to improve adhesion.
Additional Information
Connect to app to save ECG
1. Bluetooth & Location service
Check if the Bluetooth and location service functions are activated
to connect the device.
Bluetooth
&
Location
2. Connecting with the smartphone app (AT-Note)
After attaching the device, connect the device with the smartphone
app (AT-Note) to check the ECG signal.
checked to start signal storage.
Turning off the power
Once the initial one-time connection is confirmed, device
is not terminated for the duration of use.
- It is impossible to physically turn it off. When the operation time
elapses, the button color turns red and the power turns off
automatically after 3 beeps.
* ATP-C130 : 14 days
- The working status of the device can only be confirmed through the
App's real-time live Viewer.
02
Attach and Installation
The ECG signal must be
7