ATSI Handheld Inductive Loop Tester HILT 9000 Manuale d'uso - Pagina 7

Sfoglia online o scarica il pdf Manuale d'uso per Tester di circuito ATSI Handheld Inductive Loop Tester HILT 9000. ATSI Handheld Inductive Loop Tester HILT 9000 17. Handheld inductive loop tester

ATSI Handheld Inductive Loop Tester HILT 9000 Manuale d'uso
6
The LOOP test can measure a maxi-
mum Q of 15. If Q is greater than 15
the tester displays Q>15. To measure
the actual value, press the Q button on
the keypad to run a separate Q test.
The default frequency at which the
LOOP and Q tests run is 50 KHz. It can
be changed to any value within the
range from 20 KHz to 80 KHz with 1
KHz increment.
Quality (Q)
• Disconnect the detector from the loop before running
this test.
• DO NOT switch to a different test while the measure-
ments are being generated for the current test. Only
after the results of the current test are displayed can
a different test be performed.
Connect the test leads to the loop. This test requires several
seconds to run during which the tester will measure Q at
some arbitrary frequency and calculate Q for a user-selected
frequency. If a user-selected frequency is not entered prior to
pressing Q, the default 50KHz setting
will be used. Refer to the Set
Frequency section on page 7 for
full instructions to enter a different
frequency.
Both Q results will be presented on the
display as shown in Figure 4. The
calculated value is shown on the
second line and indicated with an *.
ATTENTION
www.atsi-tester.com
R=12.3
Q=12.1
L=89.2µH
Figure 3
Q=12.1
@40.5KHz
Q=14.9 @50.0KHz
*
Figure 4
50KHz