Thanks for using our products, please read this manual thoroughly before
operation.
OVERVIEW
MODEL:
TPR3003T/TPR3005T
including both stable voltage and stable current function,
circuit protection, thermal protection with load ability etc
laboratory, electronics, communications equipment maintenance, Product line, scientific research
and teaching units. If power working long hours running state, overheating will shut off the output,
when machine temperature returned to normal, output power restore default state. The machine
has perfect protection function, can protect user equipment are not consumed.
SPECIFICATION
1. Rated working conditions and external dimension
Input AC voltage
AC 220V 5%, Frequency 50Hz
(If according to the requirement can be designed in 110V 5% 60Hz)
The working conditions
Temperature rang: 0 ~40
Relative humidity less than 80%
Storage conditions
Temperature rang: -20 ~80
Relative humidity less than 80%
Shape dimension
(L)270 (W)125 (H)170mm
Accessories
users manual 1pc
power cable
1pc
2. Technical specification
Output voltage from 0 to 30V between nominal continuous adjustable.
Output current from 0 to nominal value between continuous adjustable.
Power regulation
CV less than 0.01 +1mV
CV less than 0.2 +1mV
regulated DC power supply is a high reliability,
Restricted current lower voltage, short
The power supply is suitable for the
2
Load regulation: CV less than 0.01 +5mV
CC less than 0.2 +5mA
CV less than 0.01 +7mV
CC less than 0.2 +7mA
Ripple and noise: CV 0.5mV
RMS
CC 3mA
RMS
(5Hz-1MHz)
CV 1mV
RMS
CC 5mA
RMS
Voltage accuracy:
0.5%rdg+2byte
Current accuracy:
0.5%rdg+2byte
Display resolution: 0.5%rdg+2byte
High voltage test:
Leakage current I 1 mA (Test conditions: 1700VAC/2s From the input
power to the ground, Input high AC voltage: 1700V, Time: 2 seconds).
Insulation resistance:
Insulation resistance more than 100M
From the input power to the ground, Input high DC voltage: 500Vdc, Time: 5 seconds).
INSTRUCTIONS
1. Note
(1) AC input
AC input should be 220V 10%, 50Hz
(If you are in the box 110V 10% 60Hz behind can indicate)
(2) Scattered heat
Do not exceed 40 degrees Celsius temperature environment in the local use, fan in the posterior
instrumentation, should have enough space, so scattered heat.
(3) Output voltage overshoot
When switching power supply device, the output voltage is not more than preset value.
(4) Overheating protection
This series of power is the perfect overheating protection function, can work under state
protection power will not lead to internal components overheating and burned power and user
equipment. If power transformer, internal temperature overheating will automatically cut off the
input voltage, When the internal temperature of power to the normal range, power will restore
default working state.
2. limiting Current Settings
(1)The decision to power instrument maximum safety.
(2)
Rotating voltage control knob, output voltage regulating 3V to 5V
Using a wire temporarily voltage output terminals (+) and (-) terminal circuit connection.
(3)
Rotating voltage control knob, until the CC designator light.
(4)
Adjust current knob to the current needs
I
3A
load
I
3A
load
I
3A
load
I
3A
load
I
3A
load
I
3A
load
I
3A
load
I
3A
load
(Test conditions: 500V /5s
3
dc