2B Technologies 106 Manuale di installazione - Pagina 7
Sfoglia online o scarica il pdf Manuale di installazione per Involucro 2B Technologies 106. 2B Technologies 106 14. Industrial enclosures for ozone monitors
1.
INTRODUCTION
This manual covers the Industrial enclosure offered for the 2B Technologies Series 106
Ozone Monitors. This polyester enclosure provides protection for the Ozone Monitor
while permitting easy access for connections, maintenance, and repairs. The shatter-proof
window provides for viewing the instrument display and status lights through the closed
door. A 1/4" inch compression fitting is provided for the sample inlet and outlet on all the
Series 106 Ozone Monitors. Multichannel inlet (3 or 6 channels) are offered as options.
This enclosure is not water resistant.
The Industrial enclosure is larger than the NEMA enclosure previously offered by 2B
Technologies, enabling easier access for maintenance and repairs.
2.
INSTALLATION
Please refer to the following information when installing the enclosure. For assistance,
please call 2B Technologies at (303) 273-0559 or email [email protected].
2B Technologies recommends that the Industrial Enclosure be mounted on a wall or
specially built stand. The enclosure weighs over 14 pounds with the ozone monitor
installed and requires a substantial mounting structure. The floor of the box should be
level when the box is mounted.
dimensions for the Industrial Enclosure.
2.1.
Mounting
The back of the enclosure has three holes at the top and three holes at the bottom for
mounting bolts (Figure 1). The holes are 1/4" in diameter and should be used with
stainless steel bolts and washers.
2.2.
Power
The Industrial units are set up for AC
power via connection to an AC/DC
converter mounted on the baseplate
(red square in drawing). These units
are
intended
to
electricians using conduit to bring
electrical power to the unit.
conduit connector, supplied by the user,
should be installed on the enclosure
through the hole provided in the
enclosure
bottom
drawing). Alternatively, the customer can use a 12VDC power supply and bring the
power cable into the box through either a conduit fitting or a 1/2" cable gland. The
power requirement is 100-240 VAC, 50/60 Hz, 11-28 V DC, nominally 500 mA at 12 V
DC, 6 watt.
The figures in this manual provide mechanical
be
installed
by
A 1/2"
(red
circle
in
106 Industrial Installation Manual Rev. A-1
1