3.1. BATTERY LIFE
3.1.1. The test set battery should be replaced when its load voltage has dropped to 7 volts or the
display reads LB, LO BAT or ∆ when the test set is turned on.
3.2. BATTERY REPLACEMENT
3.2.1. Any 9 Volt battery may be used but longer service will be obtained by the use of premium
batteries such as NEDA Type 1604A. The battery is accessible in the Test Set by removing the bat-
tery access door on the bottom rear of the case and dropping out the old battery. A snap type con-
nector facilitates change of battery and prevents reversal of polarity, but as an additional precau-
tion, always turn the test set off before making the battery change.
3.3. TEST SET CALIBRATION
3.3.1. The TS500 Test Set is calibrated at the factory during manufacture. Under normal operat-
ing conditions the test set will not require calibration for a period of 15 months. The calibration
due date is shown on the back of the test set.
3.3.2. For calibration procedures, please contact Dukane Seacom, Inc. or refer to the calibration
procedures in the CT200 manual.
SECTION III
MAINTENANCE
DOCUMENT NO. 03-TM-0082 REV D
Page 5 of 6