4. End Cap Assembly
4.1
Lift up module
Open cardboard box, lift module out of it and place on the platform or shelf, keep two ends of module above the
ground.
4.2
Place liquid holder
Place two liquid collection basin beneath the module ends, Figure 2.
Cautions:
Each non-anti-freezing module (GMID 12091655,12091656) contains 4 L storage solution, the major ingredients
of the liquid are 1.3% Sodium Metabisulfite, 2.5% Sodium Citric, with the pH at 4~5, CODCr at around
26,000mg/L. Each anti-freezing module (GMID 12091659, 12091658) contains 23L storage solution, the major
ingredients of the liquid are 1.2% Sodium Metabisulfite, 2.5% Sodium Citric and 60% glycerin, with the pH at 4~5,
CODCr at around 730,000mg/L. Please choose liquid collection basin with proper volume.
Please dispose the storage solution in compliance with local regulation.
4.3
Discharge storage solution
Dismount module shipping plugs (6, please refer to attachments for all the parts name and number mentioned in this
document) at two ends, and the concentrate plug (10). Remove X-rings (7) and concentrate seal (9), keep them clean and
restore for use. Gradually lift up module at filtrate end, keep the module at the inclined angel (<30 degree) and empty the
storage solution from feed opening.
Cautions:
Always keep the X-rings and seals clean, prevent them from dust and particle contamination.
Page 6 of 13
Figure 2 Liquid collection basin beneath the module ends
End Cap Assembly
Form No. 45-D04608-en, Rev 0
September 2023