COMPARE FUNCTION

When Compare function is selected, the Meter beeps and won't display "PASS",
when the measuring data is higher than the upper limited value or lower than the
lower limited value.
Pressing the COMP button will activate the Compare
function and the Meter will display "COMP" (see
image right).
Press COMP to view the upper and lower limited value.
The Meter displays "COMP" and either "HI" or "LO"
on the secondary screen (see image right).

TIMER FUNCTION

The lock function will be invalid and the Meter will start the Insulation test when
pressing the MEASURE button. When the time is due, the test is stopped.
In 'Timer' mode, the current voltage and time is shown
on the secondary display.
The Meter will display the test voltage when it is under
the insulation test.
To check the time, press
Press SELECT to check the preset time.
Press ENTER and the Meter displays "SEt" (see right).
The time can be adjusted by pressing the
Press ENTER again to confirm the adjustment.

MAX/MIN/AVG

In 'Test' mode, press the SELECT button to view max, min and average value.
In 'Hold' mode, press the SELECT button to review max, min and average value.
DAR & PI
Sometimes an insulation part with obvious drawbacks e.g. the insulation part is
broken through under high voltage, is nevertheless good with absorption ratio.
Therefore, absorption ratio cannot be used to discover local insulation drawbacks,
other than dampness and contamination.
DAR (absorbing ratio) =
PI (polarization index) =
When the preset upper limited value is lower than
the lower limited value, COMP function is invalid and the
Meter will display "- - - -" (see image left).
The Timer function can only be activated in the
Insulation test.
Enable the Timer function by pressing TIMER.
.
The LCD screen displays "TIMER" and "PRES" is shown
on the primary display, while the secondary display shows the
preset time.
R60 Sec
R15 Sec
R10 Min
R1 Min
/ .
R10 Min = Resistance value measured 10
minutes after applying the test voltage.
R1 Min = R60 Sec = Resistance value
measured 10 minutes after applying the test
voltage.
R15 Sec = Resistance value measured 10
minutes after applying the test voltage.