Apogee SP-522 Manuale d'uso - Pagina 11

Sfoglia online o scarica il pdf Manuale d'uso per Accessories Apogee SP-522. Apogee SP-522 17. Pyranometer

Apogee SP-522 Manuale d'uso
Read only registers (function code 0x3).
Float Registers
0
1
2
3
4
5
6
7
8
9
Integer Registers
40
41
42
43
44
Read/Write registers (function codes 0x3 and 0x10).
Float Registers
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
Integer Registers
48
49
50
51
52
53
calibrated output watts
detector millivolts
Reserved for Future Use
device status
(1 means device is busy, 0 otherwise)
firmware version
calibrated output watts (shifted one decimal point to the left)
detector millivolts (shifted one decimal point to the left)
Reserved for Future Use
device status (1 means device is busy, 0 otherwise)
firmware version (shifted one decimal point to the left)
baud rate (0 = 115200, 1 = 57600, 2 = 38400, 3 = 19200, 4 = 9600, any
other number = 19200
parity (0 = none, 1 = odd, 2 = even)
baud rate (0 = 115200, 1 = 57600, 2 = 38400, 3 = 19200, 4 = 9600, any
other number = 19200)
parity (0 = none, 1 = odd, 2 = even)
slave address
model number*
serial number*
number of stopbits
multiplier*
offset*
running average
heater status
slave address
model number*
serial number*
number of stopbits