Apogee SQ Series Manuale d'uso - Pagina 11

Sfoglia online o scarica il pdf Manuale d'uso per Accessories Apogee SQ Series. Apogee SQ Series 19. Quantum sensor

OPERATION AND MEASUREMENT

Connect the sensor to a measurement device (meter, datalogger, controller) capable of measuring and displaying
or recording a millivolt signal (an input measurement range of approximately 0-250 mV is required to cover the
entire range of PPFD from the sun). In order to maximize measurement resolution and signal-to-noise ratio, the
input range of the measurement device should closely match the output range of the quantum sensor. DO NOT
connect the sensor to a power source. The sensor is self-powered and applying voltage will damage the sensor.
Wiring
White: Positive (signal from sensor)
Black: Negative (signal from sensor)
Clear: Shield/Ground