Apogee SQ-618 Manuale d'uso - Pagina 6

Sfoglia online o scarica il pdf Manuale d'uso per Accessories Apogee SQ-618. Apogee SQ-618 19. Epar sensor

Apogee SQ-618 Manuale d'uso

SPECIFICATIONS

Input Voltage
Average Max Current Draw
Calibration Factor
Calibration Uncertainty
Measurement Repeatability
Long-term Drift
(Non-stability)
Non-linearity
Field of View
Spectral Range
Directional (Cosine) Response
Azimuth Error
Tilt Error
Temperature Response
Uncertainty in Daily Total
Detector
Housing
IP Rating
Operating Environment
Dimensions
Mass (with 5 m of cable)
Cable
Calibration Traceability
Apogee Instruments SQ-600 series ePAR sensors are calibrated through side-by-side comparison to the mean of
four transfer standard sensors under a reference lamp. The transfer standard sensors are recalibrated with a
quartz halogen lamp traceable to the National Institute of Standards and Technology (NIST).
RS-485 quiescent 37 mA, active 42 mA
Custom for each sensor and stored in the firmware
± 5 % (see calibration Traceability below)
Less than 1 % (up to 4000 µmol m
Less than 1 % (up to 4000 µmol m
380 to 760 nm ± 5 nm (wavelengths where response is greater than 50 %)
± 2 % at 45° zenith angle, ± 5 % at 75° zenith angle (see Directional Response below)
-0.11 ± 0.04 % per C (see Temperature Response below)
Anodized aluminum body with acrylic diffuser
-40 to 70 C; 0 to 100 % relative humidity; can be submerged in water up to depths of 30 m
5 m of shielded, twisted-pair wire; TPR jacket (high water resistance, high UV stability,
flexibility in cold conditions); pigtail lead wires; stainless steel (316), M8 connector
SQ-618-SS
5.5 to 24 V DC
RS-232 37 mA;
Less than 2 % per year
180°
Less than 0.5 %
Less than 0.5 %
Less than 5 %
Blue-enhanced silicon photodiode
IP68
30.5 mm diameter, 37 mm height
140 g
-2
-1
s
)
-2
s
-1
)