Apogee SQX Series Manuale d'uso
Spectral Response
Mean spectral response of four SQ-100x
series quantum sensors compared to PPFD
weighting function. Spectral response
measurements were made at 10 nm
increments across a wavelength range of
350 to 800 nm in a monochromator with an
attached electric light source. Measured
spectral data from each quantum sensor
were normalized by the measured spectral
response of the monochromator/electric
light combination, which was measured
with a spectroradiometer.