Apogee Instruments SU-202-SS Manuale d'uso - Pagina 7
Sfoglia online o scarica il pdf Manuale d'uso per Accessories Apogee Instruments SU-202-SS. Apogee Instruments SU-202-SS 18. Ultraviolet sensor
Anche per Apogee Instruments SU-202-SS: Manuale d'uso (16 pagine)
Cosine Response
Cosine Response Graph
Directional, or cosine, response is
defined as the measurement error at
a specific angle of radiation incidence.
Directional error for Apogee SU-200
series UV-A sensors is approximately ±
2 % and ± 5 % at solar zenith angles of
45° and 75°, respectively.
Mean cosine response of four
Apogee UV-A sensors. Cosine
response was calculated as the
relative difference of UV-A
sensors from the mean of
replicate reference UV-A
sensors deployed outdoors.
These data are the average of
the AM and PM response.