Apogee Instruments SU-221 Manuale d'uso - Pagina 8

Sfoglia online o scarica il pdf Manuale d'uso per Accessories Apogee Instruments SU-221. Apogee Instruments SU-221 20. Ultraviolet-a sensor

Apogee Instruments SU-221 Manuale d'uso
Cosine Response
Cosine Response Graph
Directional, or cosine, response is
defined as the measurement error at
a specific angle of radiation incidence.
Directional error for Apogee SU-200
series UV-A sensors is approximately ±
2 % and ± 5 % at solar zenith angles of
45° and 75°, respectively.
Mean cosine response of four
Apogee UV-A sensors. Cosine
response was calculated as the
relative difference of UV-A
sensors from the mean of
replicate reference UV-A sensors
deployed outdoors. These data
are the average of the AM and
PM response.